【Product name】
Diagnostic Kit for Cystain C(Immunochromatographic assay)
【Package specification】
25 Tests/kit
【Intended use】
It is used to quantitatively detect the content of cystatin C in human serum, plasma and whole blood samples. It is mainly used as one of the indicators reflecting glomerular filtration rate
【Test principle】
The Cys-C Rapid Test is a one-step chromatographic sandwich immunoassay designed for the quantitative measurement of Cys-C. The Cys-C antigen in the sample was first bound with the conjugated compound of fluorescent labeled Cys-C monoclonal antibody, then moved and combined with another Cys-C monoclonal antibody fixed on the nitrocellulose membrane, and the double antibody sandwich complex was formed at the detection line of the cellulose nitrate membrane. The quantitative detection results were obtained by NIR-1000 dry fluoroimmunoassay analyser.
【Components】
Name | Quantity | Component |
Test cards | 25 | It is composed of fluorescent pad (coated with fluorescent labeled Cys-C monoclonal antibody), nitrocellulose membrane (coated with Cys-C monoclonal antibody and Goat anti mouse IgG antibody), absorbent paper and backing |
Sample diluent | 25(1.50mL/tube) | Phosphate buffer |
ID card | 1 | With specific stand curve file |
The components in different batches of kits cannot be used interchangeably.
【Storage conditions and validity】
The kit should be stored at 4℃~30℃, out of direct sunlight. It is valid for 18 months. The test card should be used within 15 minutes after unsealing under the environment of 15℃~30℃ and 20% ~ 90% relative humidity.
The production date, batch number and expiration date are shown in the outer package of the product.
【Applicable instruments】
NIR-1000 dry fluoroimmunoassay analyser produced by WWHS Biotech. Inc
【Sample requirements】
【Test procedure】
【Reference interval】
Healthy non-pregnant adults are expected to have serum Cys-C values 0.5mg/L~1.1mg/L. It is strongly recommended that each laboratory should determine its own normal and abnormal values based on population.
【Interpretation of results】
【Limitations of methods】
【Performance】
1. Limits of detection
No more than 0.40mg/L.
2. Accuracy
The relative deviation from the target value is within ±15%.
3. Precision
The within and between assay coefficient of variations are within 15%.
4. Linear range
Within the linear range (0.40 ~ 9mg/L), the linear correlation coefficient R≥0.990.
【Note】
1. This kit is only used for in vitro diagnosis.
2. The test card and sample diluent are disposable and cannot be reused.
3. Please check the integrity and validity of the kit package before use, and then open the package. When it is stored at low temperature, it should be restored to room temperature (15℃ ~ 30℃) before opening the package for use. The reagents with damaged inner package and beyond the validity period cannot be used.
4. The requirements of specimen collection and storage should be strictly observed. If the specimen is turbid, it should be centrifuged and discarded before use.
5. The used kits should be treated as potential infectious substances, and all samples, reagents and potential pollutants should be disinfected and treated according to the relevant local regulations.
Other WWHS POCT Assay List
Interpretation of results
1. This reagent is only used for auxiliary detection. If the test results are abnormal, it should be reviewed in time and judged in combination with clinical symptoms.
2. For samples with AFP concentration lower than 2.5ng/ml and higher than 200ng/ml, the detection results are reported as "< 2.5ng/ml" and "> 200ng /ml", respectively.
Limitations of methods
1. This kit is only used to detect human plasma/whole blood samples
2. Due to the limitations of immunoassay methods of antigen and antibody reaction, the results cannot be used as the only basis for clinical diagnosis, but should be evaluated with all the existing clinical and experimental data.
3. The content of triglyceride in the sample shall not exceed 15mg/ml, the content of hemoglobin shall not exceed 5mg/ml, and the content of bilirubin shall not exceed 0.5mg/ml, and the relative deviation of the test results shall not exceed ±15%.
4. When the concentration of AFP in the sample is less than 20000ng/ml, there is no hook effect.
5. HAMA effect was not produced when the concentration of human anti rat in the sample was less than 50ng/ml.
6. When RF concentration in the sample is less than 2000IU/ml, the relative deviation of the test results is within ±15%.
Storage Conditions
Store the kit between 2-30°C and ensure the kits are not frozen or stored in direct sunlight. The test is valid until the expiration date printed on the foil wrapping.
Applicable Instrument
WWHS NIR-1000 dry fluoroimmunoassay analyser
FAQ:
1. What is the MOQ?
Generally if you choose the different products, our minimum order qty. also will be different. One sample available.
2. What about the delivery time?
The goods need 7-21 work days to be made upon deposit.
3. Does your product have any warranty?
Yes, we offer 6 months limited warranty for our products.
Damage due to misuse, ill treatment and unauthorized modifications and repairs are not covered by our warranty.
4. What's your payment method?
We usually accept the payment methods include PayPal, TT, and Western Union.
50% deposit in advance and balance before shipping. Buyer can choose which payment ways that you accept.
PayPal is most useful and safe, please don’t refuse PayPal fee when you choose PayPal.
5. What's your shipping method?
We provide comprehensive shipping methods.
For small quantity orders we ship by DHL Air-Express, or EMS/TNT/UPS/FEDEX Express service, it is safe and fast.
For large quantity orders we ship by Air transportation or sea transportation, we can also ship order to buyer’s cargo agent in China.
Send your inquiry directly to us