【Product name】CA125.pdf
Carbohydrate Antigen 125(CA125)Rapid Quantitative Test(Fluorescence immunoassay)
【Package specification】
25 Tests/kit
【Components】
Name | Quantity | Component |
Test cards | 25 | It is composed of fluorescent pad (coated with fluorescent labeled CA125 monoclonal mouse antibody), nitrocellulose membrane (coated with CA125 monoclonal mouse antibody and Goat anti mouse IgG antibody), absorbent paper and backing |
Sample diluent | 25(300μL/tube) | Phosphate buffer |
ID card | 1 | With specific stand curve file |
The components in different batches of kits cannot be used interchangeably.
【Storage conditions and validity】
The kit should be stored at 4℃~30℃, out of direct sunlight. It is valid for 18 months. The test card should be used within 15 minutes after unsealing under the environment of 15℃~30℃ and 20% ~ 90% relative humidity.
The production date, batch number and expiration date are shown in the outer package of the product.
【Applicable instruments】
NIR-1000 dry fluoroimmunoassay analyser produced by WWHS Biotech. Inc.
【Procedure】
【Performance】
1. Limits of detection,No more than 20U/mL.
2. Accuracy,The relative deviation from the target value is within ±15%.
3. Precision,The within and between assay coefficient of variations are within 15%.
4.Linear range,Within the linear range (20-500U/mL), the linear correlation coefficient R≥0.990.
【Note】
1. This kit is only used for in vitro diagnosis.
2. The test card and sample diluent are disposable and cannot be reused.
3. Please check the integrity and validity of the kit package before use, and then open the package. When it is stored at low temperature, it should be restored to room temperature (15℃ ~ 30℃) before opening the package for use. The reagents with damaged inner package and beyond the validity period cannot be used.
4. The requirements of specimen collection and storage should be strictly observed. If the specimen is turbid, it should be centrifuged and discarded before use.
5. The used kits should be treated as potential infectious substances, and all samples, reagents and potential pollutants should be disinfected and treated according to the relevant local regulations.